Abstract

In this paper, we propose a semi-closed microchip which can realize the probe type manipulation in the microchip. The semi-closed microchip has,a bath in the middle of microchannel to allow the insertion of micropipettes into the microchip. The bath is sealed off by thin oil film to prevent the evaporation of the solution in the bath. The seal by the oil film can maintain even if micropipettes are inserted into the bath and pulled out from the bath. The seal by the oil film was evaluated by measuring the weight of the bath after 12/24 hours. To use the semi-closed microchip for cell culture and cell analysis, the exchange of the solution in the semi-closed microchip was demonstrated. The cell fixation with the thermo sensitive gel was also demonstrated by fabricating the ITO electrodes as heaters in the bath. Cell culture and the target cell harvesting were conducted by the semi-closed microchip. The proposed semi-closed microchip can be used for cell culture, cell analysis and cell harvesting under a microscope as the biological applications for single cell analysis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.