Abstract

The most recent trend on digital hearing aid is to increase the ratio of signal to noise by directivity or to develop noise reduction algorithm inside DSP IC chip. This paper designed, fabricated and tested a digital hearing aid directivity testing device in which a micro-mouse-like the stepping motor with a speaker rotates around an examinant. Both ears of the examinant were fixed with ITE hearing aids in order to respond to receiving sound. The experimental results were compared with those of a boundary element method program for verification. The diameter of the directivity testing device was 2 m and the micro-mouse was precisely controlled by PICBASIC micro processor.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.