Abstract

To diagnose dental pulp vitality, electric pulp tester has been widely used, which is a method to test condition of nerve. However, especially in the case of patients with trauma, nerve desensitization could temporarily occur even though nerve might be recovered by blood flow within the pulp later, which implies that blood flow in dental pulp is also an important factor for diagnosing vitality. This paper described the development of a probe that relatively measured blood flow in dental pulp using photoplethysmography (PPG). The probe emits four different wavelength light sources including three visible and an infrared light. We tested which light source detect sensitively the blood flow in dental pulp. As a result, green light had the largest peak to peak voltage and the power spectrum among different wavelengths.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.