Abstract

The paper describes some features of the LaBr3(Ce) crystal in comparison with the NaI(Tl) crystal. The instrumental spectra obtained by the spectrometric detection unit with a LaBr3(Ce) crystal in the fields of high-energy capture gamma radiation at neutron calibration facility AT-140 in the energy range from 4 MeV to 9 MeV are presented. It was shown that the energy calibration of LaBr3(Ce) – based spectrometers can be performed in the range from 30 keV to 10 MeV using high-energy capture gamma radiation at AT-140 and the lanthanum intrinsic radioactivity line without resorting to OSGI sources.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.