Abstract

The authors have designed, manufactured and tested a coordinate-sensitive detector for charged particle spectroscopy. The detector can be used in the devices for the elemental analysis of materials, providing simultaneous analysis of all the elemental composition with high sensitivity and precision. The designed device is based on an integrated circuit (IC) and a microchannel plate (MCP) electron multiplier. The IC is mounted on a ceramic substrate. Ions fall on the MCP mounted above the IC. Giving rise to a pulse which typically exceeds 106 electrons, each ion falls on the detector electrodes and these pulses are counted. In this research, a two stage stack of MCPs (Hamamatsu) was used. The MCPs have a channel diameter of 12 mm on a 15 mm pitch. The results of tests carried out in a mass spectrometer are presented. The designed detector is small, light, and low-power.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.