Abstract

In this paper, we studied additional filter and Ion chamber combinations to reduce patient dose without decreasing image quality in digital chest x-ray projection. The experiment set 125 kVp, 320 mA, AEC mode. Ion chamber sensors was divided by 4 cases of combinations, then, we measured patient dose and calculated organ dose using PCXMC. Also, physical image assessment using MTF was performed. As a results, The surface entrance dose and organ dose were the lowest when selecting both left and right Ion chamber sensors under the same conditions of additional filter. In image quality assessment, The spatial frequency scored 2.494 lp/mm which was highest when selecting both right and left Ion-chambers and 0.1 mmCu filter. And to conclude, to select both right and left Ion chamber sensors and 0.1 mmCu filter will help for acquiring good quality image as well as reducing patient dose.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.