We developed a flow distributor for high-performance chromatographic processes during the purification of biopharmaceuticals. Twelve distributor candidates are provided, including four types of distributors with different-shaped holes, and three sub-types with different number of holes. We evaluated the flow distributor candidates by spreading bromophenol blue dye on the membrane , and selected the most suitable distributor Arc 45 deg. type. Using 10 ml columns with three different length-to-diamter (L/D) ratios (4.33, 0.92, and 0.34) of column, we measured peak asymmetry (As) values. The 10 ml column with 0.92 L/D ratio showed best As value. We also carried out flow experiments on the three sub-types of Arc 45 deg. distributors on the 0.92 L/D ratio 50 ml column and Arc 45 deg. 1:8 distributor showed best performance. The As values of this optimal flow distributor are 1.2 or less. In summary, this study provides optimal distributor values for chromatographic purification of biopharmaceuticals.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.