It is very important to achieve a high reproducibility in the ultrasonic measurement of bone mineral density. In this study, we examined number of sampling waveform, control of temperature, diameter of region of interest as factors to improve reproducibility. We decided the optimal number of waveforms to be converted to frequency domain as period of 1. We have minimized the effects of variable temperature and constrained generation of micro bubble by keeping temperature within a range of 32±0.5℃ with a precise temperature controlling algorithm. We also found the optimal diameter of region of interest to be 13mm. In this paper, we demonstrated the improved reproducibility by controlling various factors affecting the ultrasonic measurement of bone mineral density.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.