Abstract

The paper presents an analytical model describing the dynamics of ion cloud, taking into account the action of space charge during a motion in ion mobility spectrometer — starting from the reaction region, where the shutter forms an ion pulse, and the further drift of the formed ion pulse towards the collector. The presented model lets to estimate the degree of influence of the space charge on possible ion losses and the resolution of ion mobility spectrometer. The effect of the space charge becomes noticeable, starting with the ion density of 106 cm–3. Comparison of the results obtained using the analytical model with the results of numerical solution of the initial equations shows that they practically coincide.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.