Abstract
The use of the dosimetry have been increasingly recognized as high radiation energy and radiation treatment planning(RTP) have rapidly developed in radiotherapy. There are many types of detectors for the dosimetry such as ionization chamber, film, TLD, diode, and etc. Among such detectors, the diode detector uses a photoconductor materials that generate electrical signals by the incident radiation energy. Though many research groups are recently interested in such materials, there is few experimental results except for silicon in the radiation therapy field. In this study, the feasibility of photoconductor materials was verified as a dosimeter through the evaluation of response properties at a high radiation energy. For the fabricated detectors based on HgI2 and PbI2, reproducibility, linearity, and pulse-rate response were analyzed. Such evaluations are essential factors for the use of dosimeter. From results, linearity and reproducibility of the fabricated HgI2 detector indicated about 7% error. The fabricated PbI2 detector showed 1.7% error in linearity, and 12.2% error in reproducibility.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.