Abstract

The water supplied from dental unit water systems (DUWS) in dentistry may be heavily contaminated with bacteria and thus may be a potential source of infection for both practice staff and patients. The aim of this study was to evaluate the level of heterotrophic bacteria and to confirm the presence of opportunistic pathogens from DUWS in student clinical simulation laboratory of college of dentistry. Water samples were collected from 36 ultrasonic scalers in student clinical simulation laboratory. The levels of heterotrophic bacteria in water samples were quantified by counting colony forming units (CFUs) on R2A agar media. In addition, opportunistic pathogens were detected by using the polymerase chain reaction (PCR) method. The mean CFUs were 16,095 CFU/ml for water samples and all of water samples exceeded current American Dental Association recommendations of 200 CFU/ml. Pseudomonas species and non-tuberculous Mycobacterium species were detected in the one sample and two samples, respectively, among the 36 water samples by the PCR with specific primers for these bacteria. Our study indicated that DUWS in student clinical simulation laboratory can cause potential infection in students and participants. This study suggested the dental unit water line management and wearing personal protective equipment in student clinical simulation laboratory will be needed to reduce bacterial contamination.