Abstract

In this study, an insole-type ground reaction force (GRF) measurement system using a load cell was manufactured and configured as a system that can measure joint angle and GRF, when walking in conjunction with a commercialized inertial sensor. The data acquisition device was used to acquire synchronized data, between the inertial measurement unit (IMU) sensor and the load cell insole. A three-dimensional motion analysis system comprising six infrared cameras and two ground reaction forces, was used to check the accuracy of the gait measurement system, comprising an inertial sensor and a load cell insole. The motion and force data were acquired while performing five times six-meter walking test by five young adult male subjects (Age: 26.0±1.8, Height: 171.4±6.8 cm, Weight: 62.2±10.8 kg). It was measured and as a result of comparing the calculated sagittal joint angle with the vertical GRF, the sagittal lower extremity joint angle correlation coefficient (Pearson’s r) was 0.40 to 0.94, and the vertical GRF to be 0.98 to 0.99. It is necessary to upgrade the joint angle calculation algorithm through future research. Additionally, the possibility of clinical application for actual stroke patients will be reviewed.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call