Abstract

Purpose: Combination of total body irradiation (TBI) with chemotherapy is widely used technique for conditioning before hematopoietic stem cell transplantation for patient with hematological malignancies worldwide. Total body irradiation for patients with high height has to be divided into two parts: irradiation of upper part of the patient’s body (including head, body and part of legs) and irradiation of lower part of the patient’s body (including leg). There is an area in which the fields overlap each other – the junction area. The aim of this work is the development and verification of simple junction technique that would provide the dose distribution in the junction area from 90 to 125 % of prescribed dose.
 Material and methods: Total body irradiation was performed on the Tomotherapy machine using helical geometry of the beam delivery. Distribution of the dose in junction area was investigated. Simple solution was proposed: during the optimization of the radiotherapy plan certain margin should be maintained between upper and lower targets while dose distribution in junction area satisfies the uniformity requirements for the given irradiation geometry. The dimension of the margin was determined experimentally using a CheesePhantom and radiochromic EBT-2 films. The uniformity of dose distribution in the junction area was monitored by in vivo measurements using radiochromic EBT-2 films located on the skin surface of patients.
 Results: The dimension of the margin at which the dose in the junction area is within the range of 90 to 125 % of the prescribed dose was determined experimentally and amounted to 5.25 cm. The values of the measured dose were in the range from 97 to 105 %. In total 18 in vivo measurements of the junction area were performed. According to the results of in vivo dosimetry, the values of the doses measured in the junction area were in the range from 93 ± 3 % to 108 ± 4 %.
 Conclusion:The developed planning method with the selected plan geometry ensures satisfactory heterogeneity of the dose distribution in the area of field junction between the upper and lower irradiation regions, despite of the existing uncertainty of patient positioning. Results were confirmed by in vivo measurements. The obtained data can be used for total body irradiation of the patients using Helical Tomotherapy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.