Abstract

In this paper, we introduce the multiple frequency bio-electrical impedance analysis method for body composition analysis. And then we implement the multiple frequency bio-electrical impedance analysis system. Overall system consists of: multiple frequency alternating current signal generator contained alternating current signal, phase signal detector, voltage signal detector, micro controller, in-out device(key-pad LCD), conductivity electrodes, system power. We explain the architecture of the system and required theory to implement the system. In order to investigate the clinical significance of the body composition data, compare to the data measured by the expert body composition analyzer which provide high reproduction and precision. Finally, experimental results which are the correlation between the measured data show the very high reproduction performance of the body composition analysis in the proposed system.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call