Notwithstanding the extensive use of conventional compression devices in managing venous disorders, these modalities have shortages that diminish their treatment efficacy and lessen patient adherence to therapy. The purpose of this study was to develop an improved compression system that eliminates the flaws of the existing devices. A motorized bandage was designed that takes advantage of continuous feedback from force-sensing resistors to apply reproducible, controlled pressure on the lower extremities. The performance of the device in enhancing venous return was explored in a pilot test on 11 healthy participants, wherein graded lower body negative pressure was employed as a surrogate of passive standing. Each subject underwent two experiments; with and without pressure application over the calves. A two-way repeated-measures analysis of variance revealed a significant difference in the mean hemodynamic responses when the compression bandage was in action (p < .05). Specifically, a meaningful increase was observed in mean arterial pressure by 5%, diastolic blood pressure by 8% and left ventricular ejection time by 4%; and a significant decrease of 5% and 6% was noticed in heart rate and pulse pressure, respectively. These results demonstrate the capability of the designed system in attenuating the imposed orthostatic stress on cardiovascular system.