The current study aims to compare the split x-jaw planning technique of volumetric modulated arc radiotherapy (VMAT) with the traditional open and limited jaw techniques of VAMT in nasopharyngeal carcinoma treatment. The multi-leaf collimators on the varian linear accelerator move on a carriage with a maximum leaf span of 15 cm. Therefore, treatment of larger planning target volumes, such as in nasopharyngeal cancer with traditional open and limited jaw technique, yields compromised dose distribution. Computed tomography data sets of 10 nasopharynx cancer patients were enrolled for the study. For each case, three separate treatment plans were generated viz. open, limited, and split x-jaw planning techniques with similar planning objectives. Only PTVs requiring a field size larger than 18 cm in the x-jaw position were considered. Comparable results were obtained regarding organs at risk (OAR) sparing in all the techniques. The target dose coverage with split x-jaw VMAT was superior to both open and limited jaw planning techniques, with a statistically significant difference in the intermediate dose planning target volumes (PTVs) (PTV59.4), P < 0.05. However, the split technique's dose to the spinal cord and larynx was significantly lower (P < 0.05). The split x-jaw planning technique of VMAT can be adapted for larger PTVs requiring an x-jaw of more than 15 cm. The only concern with this technique is the increased MU.