Intra-dialytic hypotension (IDH) remains the commonest problem associated with routine haemodialysis treatments. Fluid shifts from intracellular(ICW) and extracellular(ECW) compartments to refill plasma volume during haemodialysis with ultrafiltration. We studied the effect of relative changes in ICW and ECW indifferent body segments using multifrequency segmental bioimpedance during haemodialysis and IDH episodes. Of 42 haemodialysis patients,16 patients (38.1%) developed IDH within the first hour of dialysis. Patients with and without early IDH were well-matched for demographics and starting bioimpedance measurements. However, after 60 min, the relative change in in ECW/ICW ratio between the non-fistula arm and leg was significantly different for the early IDH group median -1.07 (-3.33 to 0.8) versus 0.61 (-0.78 to 1.8), p < 0.05, whereas there no differences in ultrafiltration rate, relative blood volume monitoring or on-line clearance. Monitoring serial changes in fluid status in different body compartments with bioimpedance may potentially prevent IDH in the future.