We evaluated whether improved increase delivery efficiency of weight optimized dynamic conformal arc (WO-DCA) therapy in comparison to volumetric modulated arc therapy (VMAT) with single isocenter for SRS treatment of very small volume and multiple brain metastases (BMs). 20 patients having a less than 1 cc volume and 2≤, ≥4 of multiple BMs, redesigned for 20 Gy in 1 fraction using WO-DCA and VMAT techniques with double full coplanar and three partial noncoplanar arcs. Plan qualities were compared using tumor coverage, conformity index (CI), gradient index (GI), V4Gy, V10Gy, and V12Gy volumes of brain, monitor units (MUs), and percent of quality assurance pass rate (QA%). Both techniques satisfied clinical requirements in coverage and CI. VMAT had a significantly higher MU and mean GI than WO-DCA (for MUs; 2330 vs. 1991; P < 0.001, and for GI; 4.72 vs. 3.39; P < 0.001). WO-DCA was found significantly lower V4Gy (171.11 vs. 232.80 cm3, P < 0.001), V10Gy (25.82 vs. 29.71 cm3, P < 0.05), and V12Gy (14.35 vs. 17.28 cm3, P < 0.05) volumes than VMAT. WO-DCA was associated with markedly increase QA pass rates for all plans (97.65% vs. 92.64%, P < 0.001). WO-DCA may be the first choice compared to the VMAT in reducing the dose in the brain and minimizing small-field dosimetric errors for very small SRS treatment of brain metastases in the range of ≤ 1 cc and 2≤, ≥4.