The production of 51Cr-labelled ethylenediaminetetraacetic acid (51Cr-EDTA), a validated and widely used radio-isotopic tracer for glomerular filtration rate (GFR) measurement in Europe, was recently halted by the manufacturer. Technetium-99m-diethylenetriaminepentaacetic acid (99mTc-DTPA) clearance has so far mostly been restricted to assessment of separate renal function by scintigraphy, but scarcely used and validated for GFR measurement. We compared the performances of 51Cr-EDTA and 99mTc-DTPA for GFR and extracellular fluid measurement. In a multi-centre prospective study, 51Cr-EDTA and 99mTc-DTPA were simultaneously injected into 88 patients, and their urinary and plasma clearances, as well as their volumes of distribution, were measured during seven 30-min periods after a 90-min equilibrium time. Mean age was 52.2 ± 14.5years, 59% were men. Urinary clearances of 51Cr-EDTA and 99mTc-DTPA were 64.1 ± 27.6 and 66.1 ± 28.0mL/min, respectively, with a mean bias of 2.00 ± 2.25mL/min, an accuracy within 10% of 95% [95% CI 91-99], and a coefficient of determination (R2) of 0.994. Plasma clearances of 51Cr-EDTA and 99mTc-DTPA were 66.1 ± 25.8 and 68.1 ± 26.6mL/min, respectively, with a mean bias of 1.96 ± 3.32mL/min, an accuracy within 10% of 91% [95% CI 85-97] and a R2 of 0.985. Distribution volumes were 17.3 ± 4.6 L for 51Cr-EDTA and 16.6 ± 4.6 L for 99mTc-DTPA (R2 0.930). The accuracy and precision of 99mTc-DTPA clearance, compared to 51Cr-EDTA clearance, was excellent for both urinary and plasma clearance methods, despite an approximate 2mL/min overestimation, showing that the tracer is a reliable alternative to 51Cr-EDTA for GFR measurement.