BackgroundGas Chromatography Isotope Ratio Mass Spectrometry (GC-C-IRMS) has long been used in routine laboratories to determine the δ13C values of anabolic steroids in urine, differentiating between, e.g., endogenous and synthetic testosterone (T) in sports doping control. Until now, liquid chromatography (LC-IRMS) has not been used. The LC-IRMS setup doesn't allow organic solvents or modifiers in the mobile phase for δ13C determinations. Mid-to non-polar analytes such as steroids can be analysed in water heated to High Temperatures (HT, up to 200 °C) because at 200 °C has a similar polarity as 80/20 methanol/water at ambient temperature. In this work, we developed a method for steroids in urine, extending the application of the LC-IRMS to non-polar analytes in complex matrices. ResultAn HT-LC-IRMS method capable of determining the δ13C values of four steroids (i.e., testosterone (T), 5α-androstane-3α,17β-diol (ααβ), 5β-androstane-3α,17β-diol (βαβ) and pregnanetriol (PT)) in urine was developed and validated. Accuracy ranged from 0.23 ‰ (ααβ and βαβ) to 0.49 ‰ (T), and the detection limit was set at 10 ng mL−1 (T, ααβ+βαβ). The validation data and a comparison of authentic urine samples analysed with HT-LC-IRMS and GC-C-IRMS indicated a comparable performance between HT-LC-IRMS and GC-C-IRMS. SignificanceHT-LC-IRMS can be used to determine δ13C values of anabolic steroids, extending the applicability of both HT-LC and LC-IRMS to non-polar substances determined in a complex matrix in routine laboratory practice.