2,316 publications found
Sort by
True intratracheal oxygen concentration delivered by SentriO Oxy™ masks under various respiratory conditions: a bench study.

SentriO Oxy™ is a newly available, Food and Drug Administration-approved oxygenation mask system that provides high oxygenation, even on low-flow (5-10L/min) oxygen. This study aimed to accurately measure the intratracheal fraction of inspired oxygen (FiO2) using SentriO Oxy™ masks under relatively low oxygen flow rates. A manikin-ventilator-test lung simulation system was used. We measured FiO2 at the level of the carina, 5 minutes after applying 45 different respiratory parameter combinations using SentriO Oxy™ masks. Tidal volume (TV) was set to 300, 500, and 700 mL; respiratory rate (RR) was set to 8, 12, 16, 20, and 24 breaths per minute; and oxygen flow rate was set to 6, 8, and 10L/min. Our hypothesis was that FiO2 would be proportional to the difference between oxygen flow rate and minute ventilation. FiO2 measured by smaller TV, lower RR, or higher oxygen flows revealed a significantly higher value, confirming our hypothesis. In addition, using linear regression analysis, we found that TV, RR, and oxygen flow were all significant factors influencing the measured FiO2. Our experiment proposed two prediction equations considering the oxygen flow rate, TV, and RR. The results of our study may provide information and prediction of FiO2 for clinicians to use SentriO Oxy™ masks during sedative anesthetic procedures under low oxygen flow rates.

Open Access
Relevant
Comparative performance of stimpod electromyography with mechanomyography for quantitative neuromuscular blockade monitoring.

The purpose of this study was to compare the Stimpod electromyograph neuromuscular blockade monitor to mechanomyography, which is widely considered to be the reference standard. The Stimpod electromyograph was used with its designated electrode array on the same hand as the mechanomyograph. Pairs of train-of-four measurements were recorded every 0.5-2min. When the train-of-four count was zero on the electromyograph monitor, pairs of post tetanic count measurements were recorded every 2.5min, instead of train-of-four measurements. Measurements were recorded from immediately after induction of anesthesia until just before emergence. Stimulation current was set to 60 mA with a duration of 200 microsec. The mechanomyography recording system recorded each twitch waveform for analysis. High resolution electromyograph waveforms were also recorded using a datalogger accessory provided by the manufacturer, facilitating inspection of individual waveforms. The administration of neuromuscular blocking drugs was left up to the discretion of the anesthesia care team. Twenty-three patients contributed 1,088 data pairs suitable for analysis. Bland-Altman analysis of 415 pairs of train-of-four ratios showed a bias of 0.028 and limits of agreement of -0.18 and 0.24. Two hundred seventy-three train-of-four count data pairs were compared by Cohen's quadratically weighted kappa which was calculated to be 0.44, indicating moderate agreement. Three hundred thirty-eight post tetanic count data pairs were compared by Cohen's quadradically weighted kappa which was calculated to be 0.80, indicating substantial agreement. The electromyograph produced results that were comparable to the mechanomyograph.

Relevant
Accuracy and clinical utility of heart rate variability derived from a wearable heart rate monitor in patients undergoing major abdominal surgery.

Low heart rate variability (HRV) can potentially identify patients at risk of intraoperative hypotension. However, it is unclear whether cheaper, readily accessible consumer heart rate (HR) monitors can provide similar utility to clinical Holter electrocardiograph (ECG) monitors. The objectives of this study were (1) to assess the validity of using the Polar H10 HR monitor as an alternative to a clinical Holter ECG and (2) to test total power (TP) as a predictor of intraoperative hypotension. The primary outcome was the level of agreement between Polar H10 and Holter ECG. Twenty-three patients undergoing major abdominal surgery with general anesthesia had 5-minute HR recordings taken concurrently with both devices during a pre-anesthetic consultation. Agreement between Polar H10 and Holter ECG was compared via Bland-Altman analysis and Lin's Concordance Correlation Coefficient. Patients were divided into groups based on TP < [Formula: see text]and TP > [Formula: see text]. Intraoperative hypotension was defined as MAP < 60 mmHg, systolic blood pressure < 80 mmHg, or 35% decrease in MAP from baseline.There was substantial agreement between Polar H10 and Holter ECG for average R-R interval, TP and other HRV indices. Reduced TP (< 500 [Formula: see text]) had a high sensitivity (80%) and specificity (100%) in predicting intraoperative hypotension. Patients with reduced TP were significantly more likely to require vasoactive drugs to maintain blood pressure.The substantial agreement between Polar H10 and Holter ECG may justify its use clinically. The use of preoperative recordings of HRV has the potential to become part of routine preoperative assessment as a useful screening tool to predict hemodynamic instability in patients undergoing general anesthesia.

Relevant
Renal medullary oxygenation during laparoscopic vs open surgery: the impact of blood pressure management-a pilot randomized controlled trial.

The impact of blood pressure targets and surgical approach (laparoscopic or open) on continuous urinary oxygenation (PuO2), a validated surrogate of renal medullary PO2, during general surgery, is unclear. We aimed to assess the effects of different blood pressure targets and surgical procedures on PuO2. We randomized patients receiving either laparoscopic or open surgery into two mean arterial pressure (MAP) target groups: usual MAP or a high MAP. We measured PuO2 in real-time and analyzed it according to the type of surgery and blood pressure target. The study was retrospectively registered on the 5th of July 2023 (ACTRN12623000726651). We included 43 participants who underwent either laparoscopic (n = 20) or open surgery (n = 23). We found that PuO2 significantly decreased during both laparoscopic and open surgery under a usual blood pressure target (- 51% and - 49%, respectively). However, there was a sharper fall with laparoscopic surgery resulting in a higher PuO2 with open surgery (mean difference: 11 ± 1 mmHg higher; p < 0.001). Targeting a higher MAP resulted in a higher PuO2 over time during laparoscopic surgery (mean difference: 7 ± 1 mmHg, p < 0.001). In contrast, targeting a usual MAP resulted in a higher PuO2 during open surgery (mean difference: 7 ± 1 mmHg, p < 0.001). Surgical approach and intraoperative blood pressure targets significantly impact urinary oxygenation. Further studies with larger sample sizes are needed to confirm these findings and understand their potential clinical implications.Registration number: ACTRN12623000726651; Date of registration: 05/07/2023 (retrospectively registered).

Open Access
Relevant
Measurement accuracy of a microwave doppler sensor beneath the mattress as a continuous respiratory rate monitor: a method comparison study.

Non-contact continuous respiratory rate monitoring is preferred for early detection of patient deterioration. However, this technique is under development; a gold standard respiratory monitor has not been established. Therefore, this prospective observational method comparison study aimed to compare the measurement accuracy of a non-contact continuous respiratory rate monitor, a microwave Doppler sensor positioned beneath the mattress, with that of other monitors. The respiratory rate of intensive care unit patients was simultaneously measured using a microwave Doppler sensor, capnography, thoracic impedance pneumography, and a piezoelectric sensor beneath the mattress. Bias and 95% limits of agreement between the respiratory rate measured using capnography (standard reference) and that measured using the other three methods were calculated using Bland-Altman analysis for repeated measures. Clarke error grid (CEG) analysis evaluated the sensor's ability to assist in correct clinical decision-making. Eighteen participants were included, and 2,307 data points were analyzed. The bias values (95% limits of agreement) of the microwave Doppler sensor, thoracic impedance pneumography, and piezoelectric sensor were 0.2 (- 4.8 to 5.2), 1.5 (- 4.4 to 7.4), and 0.4 (- 4.0 to 4.8) breaths per minute, respectively. Clinical decisions evaluated using CEG analyses were correct 98.1% of the time for the microwave Doppler sensor, which was similar to the performance of the other devices. The microwave Doppler sensor had a small bias but relatively low precision, similar to other devices. In CEG analyses, the risk of each monitor leading to inadequate clinical decision-making was low. UMIN000038900, February 1, 2020.

Relevant
Agreement of zero-heat-flux thermometry with the oesophageal and tympanic core temperature measurement in patient receiving major surgery.

To identify and prevent perioperative hypothermia, most surgical patients require a non-invasive, accurate, convenient, and continuous core temperature method, especially for patients undergoing major surgery. This study validated the precision and accuracy of a cutaneous zero-heat-flux thermometer and its performance in detecting intraoperative hypothermia. Adults undergoing major non-cardiac surgeries with general anaesthesia were enrolled in the study. Core temperatures were measured with a zero-heat-flux thermometer, infrared tympanic membrane thermometer, and oesophagal monitoring at 15-minute intervals. Taking the average value of temperature measured in the tympanic membrane and oesophagus as a reference, we assessed the agreement using the Bland-Altman analysis and linear regression methods. Sensitivity, specificity, and predictive values of detecting hypothermia were estimated. 103 patients and one thousand sixty-eight sets of paired temperatures were analyzed. The mean difference between zero-heat-flux and the referenced measurements was -0.03 ± 0.25°C, with 95% limits of agreement (-0.52°C, 0.47°C) was narrow, with 94.5% of the differences within 0.5°C. Lin's concordance correlation coefficient was 0.90 (95%CI 0.89-0.92). The zero-heat-flux thermometry detected hypothermia with a sensitivity of 82% and a specificity of 90%. The zero-heat-flux thermometer is in good agreement with the reference core temperature based on tympanic and oesophagal temperature monitoring in patients undergoing major surgeries, and appears high performance in detecting hypothermia.

Open Access
Relevant
Electrocardiogram alterations in non-traumatic brain injury: a systematic review.

The presence of abnormal electrocardiograms in individuals without known organic heart disease is one of the most common manifestations of cardiac dysfunction occurring during acute non traumatic brain injury. The primary goal of the present review is to provide an overview of the available data and literature regarding the presence of new-onset electrocardiographic (ECG) alterations in acute non traumatic brain injury. The secondary aim is to identify the incidence of ECG alterations and consider the prognostic significance of new-onset ECG changes in this setting. To do so, English language articles from January 2000 to January 2022 were included from PubMed using the following keywords: "electrocardiogram and subarachnoid hemorrhage", "electrocardiogram and intracranial hemorrhage", "Q-T interval and subarachnoid hemorrhage ", "Q-T interval and intracranial bleeding ", "Q-T interval and intracranial hemorrhage", and "brain and heart- interaction in stroke". Of 3162 papers, 27 original trials looking at electrocardiogram alterations in acute brain injury were included following the PRISMA guideline. ECG abnormalities associated with acute brain injury could potentially predict poor patient outcomes. They could even herald the future development of neurogenic pulmonary edema (NPE), delayed cerebral ischemia (DCI), and even in-hospital death. In particular, patients with SAH are at increased risk of having severe ventricular dysrhythmias. These may contribute to a high mortality rate and to poor functional outcome at 3months. The current data on ECG QT dispersion and mortality appear less clearly associated. While some patients demonstrated poor outcomes, others showed no relationship with poor outcomes or increased in-hospital mortality. Observing ECG alterations carefully after cerebral damage is important in the critical care of these patients as it can expose preexisting myocardial disease and change prognosis.

Relevant
Video plethysmography for contactless blood pressure and heart rate measurement in perioperative care.

The purpose of this study was to evaluate the feasibility and accuracy of remote Video Plethysmography (VPPG) for contactless measurements of blood pressure (BP) and heart rate (HR) in adult surgical patients in a hospital setting. An iPad Pro was used to record a 1.5-minute facial video of the participant's face and VPPG was used to extract vital signs measurements. A standard medical device (Welch Allyn) was used for comparison to measure BP and HR. Trial registration: NCT05165381. Two-hundred-sixteen participants consented and completed the contactless BP and HR monitoring (mean age 54.1 ± 16.8 years, 58% male). The consent rate was 75% and VPPG was 99% successful in capturing BP and HR. VPPG predicted SBP, DBP, and HR with a measurement bias ± SD, -8.18 ± 16.44 mmHg, - 6.65 ± 9.59 mmHg, 0.09 ± 6.47 beats/min respectively. Pearson's correlation for all measurements between VPPG and standard medical device was significant. Correlation for SBP was moderate (0.48), DBP was weak (0.29), and HR was strong (0.85). Most patients were satisfied with the non-contact technology with an average rating of 8.7/10 and would recommend it for clinical use. VPPG was highly accurate in measuring HR, and is currently not accurate in measuring BP in surgical patients. The VPPG BP algorithm showed limitations in capturing individual variations in blood pressure, highlighting the need for further improvements to render it clinically effective across all ranges. Contactless vital signs monitoring was well-received and earned a high satisfaction score.

Relevant