Abstract

Oxytetracycline impregnated XAD-4 resin was prepared and then used for the online preconcentration of zirconium (IV) from aqueous solution. HCl was used for the elution of zirconium from the complex retained on XAD-4 resin. Arsenazo-III solution was used as chromogenic reagent for the spectrophotometric determination of eluted zirconium (IV) ions. The developed system showed greater sensitivity and selectivity as compared to many reported methods. The current study achieved a detection limit of 0.47 and 0.10 µgL−1, preconcentration factor of 52 and 247 and sample throughput of 40 and 10 samples per hour for concentration range from 10 to 60 µg L−1 (60s preconcentration time) and from 2 to 12 µgL−1 (300s preconcentration time) respectively. The method was successfully applied for the determination of Zirconium (IV) from spiked water and IAEA-soil sample.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.