Abstract

X-band room temperature spectra obtained by rapid-scan, continuous wave, field-swept echo-detected and Fourier transform electron paramagnetic resonance (FTEPR) were compared for three samples with long electron spin relaxation times: amorphous hydrogenated silicon (T1 = 11 μs, T2 = 3.3 μs), 0.2% N@C60 solid (T1 = 120–160 μs, T2 = 2.8 μs) and neutral single substitutional nitrogen centres (NS0) in diamonds (T1 = 2300 μs, T2 = 230 μs). For each technique, experimental parameters were selected to give less than 2% broadening of the lineshape. For the same data acquisition times, the signal-to-noise for the rapid-scan spectra was one-to-two orders of magnitude better than for continuous wave or field-swept echo-detected spectra. For amorphous hydrogenated silicon, T2* (∼ 10 ns) is too short to perform FTEPR. For 0.2% N@C60, the signal-to-noise ratio for rapid scan is about five times better than for FTEPR. For NS0 the signal-to-noise ratio is similar for rapid scan and FTEPR.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.