Abstract
Abstract Controlling beam position in three-dimensional pencil-beam scanning is important because errors in beam position disturb the uniformity of irradiated dose distribution. To suppress this, we have developed a simple procedure for beamline tuning. Beamline tuning is performed with steering magnets and fluorescent screen monitors to converge a beam’s trajectory to a central orbit. The misalignment between beam position and a reference axis is determined through a verification system, which consists of a screen monitor and an acrylic phantom. If beam position deviates from the reference axis, two pairs of steering magnets, which are placed downstream of the beam transport line, correct the beam. These adjustments are iterated until the deviations for 11 energies of the beam are within 0.5 mm of the reference axis. To demonstrate the success of the proposed procedure, we used it to perform beam commissioning at the Kanagawa Cancer Center.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.