We demonstrate a novel hybrid-beam scanning-based quantitative optical coherence tomography angiography (OCTA) that provides high-speed wide dynamic range blood flow speed imaging. The hybrid-beam scanning scheme enables multiple OCTA image acquisitions with a wide range of multiple time intervals simultaneously providing wide dynamic range blood flow speed imaging independent of the blood vessel orientation, which was quantified over a speed range of 0.6∼104 mm/s through the blood flow phantom experiments. A fully automated high-speed hybrid-beam scanning-based quantitative OCTA system demonstrates visualization of blood flow speeds in various vessels from the main arteries to capillaries in a 4 mm×4 mm area (1024 A-lines × 512 B-scans) in vivo in 20s, showing its potential as a useful imaging tool for various biomedical applications.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.