Abstract
A 5-pass KTa1−xNbxO3 (KTN) deflector based on a DC charge technique for the swept light source of an optical coherence tomography system is proposed. The DC charge technique used in a 3-pass KTN deflector enables us to obtain the long-term stability of the optical power without any degradation of the important features of the swept light source. Experimental results confirm that the coherence length of a swept light source with the 5-pass KTN deflector using the DC charge technique is almost equal to that using a precharge technique reported previously. This means that the DC charge technique does not degrade the coherence length. There are limitation values for applied voltage, the KTN electrode gap, relative dielectric constant, and sweep frequency of the KTN deflector in terms of obtaining long coherence length. We believe that a 7-pass KTN deflector is effective for obtaining a longer coherence length.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.