Abstract

The design, realization and verification of a system for personalized rehabilitation of Spinal Cord Injury (SCI) patients are here presented. It consists of two main parts: (i) one induces in the patient the physiotherapic movement proposed by the trainer with Functional Electrical Stimulation (FES) and (ii) the other integrates a muscle fatigue feedback evaluator based on acustic principle by wearable microphone. The importance of this fatigue feedback is crucial since SCI patients are unable to feel muscle soreness, which is an important parameter in FES rehabilitation. Indeed, the efficacy of this treatment is limited by the capacity of the muscle fibers to carry out the induced task. To optimize the feedback module, the performance of two different condenser microphones (analogic and digital) were compared each other as well as with the traditional method based on Electromyogram (EMG). Among these techniques, it was found that the digital microphone provides the best performance in fatigue estimation over time.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.