Abstract

Development of pre-clinical testing methodologies is an important goal for improving prediction of artificial replacement joint performance and for guiding future device design. Total disc replacement wear and the potential for osteolysis is a growing concern, therefore a parametric study on the effects on wear of altered kinematics and loading was undertaken. A standard ISO testing protocol was modified in order to study the wear behaviour of lumbar total disc replacements when subject to low cross shear input kinematics, reduced axial loading and smaller flexion–extension magnitude. Volumetric wear, bearing surface topography, and wear debris biological reactivity were assessed. The ISO standard results were expected, however, the very low cross shear test produced a level of wear approximately two orders of magnitude higher than that reported for zero cross shear motions on UHMWPE bearings. When the osteolytic potential of the wear particles was calculated, all total disc replacement simulations had lower predicted osteolytic potential compared to total hip replacements, as a consequence of the generally lower wear rates found.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.