Abstract

We present a practical calibration method of the detection efficiency (DE) of single photon detectors (SPDs) in a wide wavelength range from 480 nm to 840 nm. The setup consists of a GaN laser diode emitting a broadband luminescence, a tunable bandpass filter, a beam splitter, and a switched integrating amplifier which can measure the photocurrent down to the 100 fA level. The SPD under test with a fibre-coupled beam input is directly compared with a reference photodiode without using any calibrated attenuator. The relative standard uncertainty of the DE of the SPD is evaluated to be from 0.8% to 2.2% varying with wavelength (k = 1).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.