Abstract

ABSTRACTA novel water-based condensation particle counter has been developed using a patented, single-flow mixing (SFM) condenser that permits a conventional thermal approach of using a hot saturator followed by a cold condenser to activate and grow particles for counting with an optical detector. A computational fluid dynamics (CFD) model of the internal flow, temperature, and vapor profiles was used to predict the effectiveness of the SFM condenser. Using the results from the CFD model, the counting efficiency was numerically calculated for pure water droplets, and the CPC cut-point (i.e., 50% counting efficiency) was predicted to be 8.3 nm. The experimental performance of the new CPC was measured with differential mobility analyzer-classified, monodisperse particles. The measured cut-points were 8.2 nm for Ag particles and 3.9 nm for NaCl particles. The reduction in the cut-point for NaCl is the result of a compound effect: water uptake by NaCl particles, which increases their size before entering into...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.