Abstract

Mescher-Garwood (MEGA) editing with spin echo full intensity acquired localization (MEGA-SPECIAL, MSpc) is a technique to acquire γ-aminobutyric acid (GABA) without macromolecule (MM) contamination at a TE of 68 ms. However, due to the requirement of multiple shot-localization, it is often susceptible to subject motion and B0 inhomogeneity. A method is presented for real-time shim and motion correction (ShMoCo) using volumetric navigators to correct for motion and motion-related B0 inhomogeneity during MSpc acquisition. A phantom experiment demonstrates that ShMoCo restores the GABA peak and improves spectral quality in the presence of motion and zero- and first-order shim changes. The ShMoCo scans were validated in three subjects who performed up-down and left-right head rotations. Qualitative assessment of these scans indicates effective reduction of subtraction artefacts and well edited GABA peaks, while quantitative analysis indicates superior fitting and spectral quality relative to scans with no correction. Copyright © 2015 John Wiley & Sons, Ltd.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.