Abstract

To visualize the X-ray fields, overlaps, and over-beaming on the skin surface during spiral head CT scanning. The measured pitch factors were determined by measuring 3 rotation times, 11 table-feed speeds, and an X-ray beam width. The X-ray fields, overlaps, and over-beaming on the skin surface were calculated via computer-aided design-based X-ray beam modeling, and the values obtained using the nominal pitch and measured pitch factors were compared. The X-ray fields with measured pitch factors exceeded those with nominal pitch factors. The overlaps increased with a decrease in the nominal pitch and measured pitch factors and were observed even at a nominal pitch factor of 1.0. The most stretched over-beaming field was observed with the measured pitch factor of 0.670. The technique can show the overlaps of the X-ray fields and may determine the adequate start angle to prevent overlaps to the eye lens.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.