Visualisation of lenticulostriate arteries using photon counting detector CT angiography: A retrospective cohort study.

  • Abstract
  • Literature Map
  • Similar Papers
Abstract
Translate article icon Translate Article Star icon
Take notes icon Take Notes

Visualisation of lenticulostriate arteries using photon counting detector CT angiography: A retrospective cohort study.

Similar Papers
  • Research Article
  • Cite Count Icon 47
  • 10.1097/rli.0000000000000901
Spectral Shaping Via Tin Prefiltration in Ultra-High-Resolution Photon-Counting and Energy-Integrating Detector CT of the Temporal Bone.
  • Jun 24, 2022
  • Investigative Radiology
  • Jan-Peter Grunz + 7 more

Hardening the x-ray beam, tin prefiltration is established for imaging of high-contrast subjects in energy-integrating detector computed tomography (EID-CT). With this work, we aimed to investigate the dose-saving potential of spectral shaping via tin prefiltration in photon-counting detector CT (PCD-CT) of the temporal bone. Deploying dose-matched scan protocols with and without tin prefiltration on a PCD-CT and EID-CT system (low-/intermediate-/full-dose: 4.8/7.6-7.7/27.0-27.1 mGy), 12 ultra-high-resolution examinations were performed on each of 5 cadaveric heads. While 120 kVp was applied for standard imaging, the protocols with spectral shaping used the highest potential available with tin prefiltration (EID-CT: Sn 150 kVp, PCD-CT: Sn 140 kVp). Contrast-to-noise ratios and dose-saving potential by spectral shaping were computed for each scanner. Three radiologists independently assessed the image quality of each examination with the intraclass correlation coefficient being computed to measure interrater agreement. Regardless of tin prefiltration, PCD-CT with low (171.2 ± 10.3 HU) and intermediate radiation dose (134.7 ± 4.5 HU) provided less image noise than full-dose EID-CT (177.0 ± 14.2 HU; P < 0.001). Targeting matched image noise to 120 kVp EID-CT, mean dose reduction of 79.3% ± 3.9% could be realized in 120 kVp PCD-CT. Subjective image quality of PCD-CT was better than of EID-CT on each dose level ( P < 0.050). While no distinction was found between dose-matched PCD-CT with and without tin prefiltration ( P ≥ 0.928), Sn 150 kVp EID-CT provided better image quality than 120 kVp EID-CT at high and intermediate dose levels ( P > 0.050). The majority of low-dose EID-CT examinations were considered not diagnostic, whereas PCD-CT scans of the same dose level received satisfactory or better ratings. Interrater reliability was excellent (intraclass correlation coefficient 0.903). PCD-CT provides superior image quality and significant dose savings compared with EID-CT for ultra-high-resolution examinations of the temporal bone. Aiming for matched image noise, high-voltage scan protocols with tin prefiltration facilitate additional dose saving in EID-CT, whereas superior inherent denoising decreases the dose reduction potential of spectral shaping in PCD-CT.

  • Research Article
  • Cite Count Icon 1
  • 10.1007/s00330-024-11204-6
Photon-counting detector CT provides superior subsolid nodule characterization compared to same-day energy-integrating detector CT.
  • Nov 28, 2024
  • European radiology
  • Jinhua Wang + 16 more

To investigate the image quality and the performance of photon-counting detector (PCD) CT compared to conventional energy-integrating detector (EID) CT in identifying subsolid nodule (SSN) characteristics. Participants with SSNs who underwent same-day EID CT and PCD CT between October 2023 and April 2024 were prospectively included. The 1.0 mm EID CT images and, subsequently, 1.0 mm, 0.4 mm, and 0.2 mm PCD CT images were reviewed to assess image noise and subjective image quality on a 5-point Likert scale. SSN characteristics, including lobulation, spiculation, pleural retraction, air cavities, intra-nodular vessel signs, internal vascular changes, and heterogeneous solid components, were evaluated. Additionally, a step-by-step observation and comparison method was used to determine the presence of any additional characteristics. Forty-eight participants (mean age: 56 ± 11 years; 16 males) with 89 SSNs were included. PCD CT significantly reduced radiation dose when using matched scans (1.79 ± 0.39 vs 2.17 ± 0.57 mSv, p < 0.001). Compared to 1.0 mm EID CT, 1.0 mm PCD CT images exhibited significantly lower objective image noise and higher subjective image quality (all p < 0.001). Compared to EID CT, PCD CT demonstrated enhanced visualization of subtle characteristics, except for lobulation, with a 0.4 mm section thickness offering a favorable balance between ultra-high resolution and perceived image quality for radiologists. PCD CT facilitated radiation dose reduction and outperformed conventional EID CT in terms of image quality and visualization of SSN characteristics. Question PCD CT, featuring ultra-high-resolution mode acquisition and a thinner reconstruction, has not been fully explored for characterizing SSNs. Findings Compared to EID CT, PCD CT was associated with lower objective image noise, higher subjective image quality, and superior SSN characterization. Clinical relevance PCD CT effectively reduced the radiation dose delivered to the patients and enabled more precise SSN characterization.

  • Research Article
  • Cite Count Icon 1
  • 10.1007/s00247-025-06336-w
Comparative radiation dose analysis in pediatric high-pitch cardiac CTA using photon-counting versus energy-integrating detector CT.
  • Nov 1, 2025
  • Pediatric radiology
  • Gladys M Arguello Fletes + 5 more

Previous studies have shown improved image quality in pediatric cardiac imaging using photon-counting detector CT (PCDCT). However, these studies did not evaluate image quality and radiation dose when utilizing the full spectral capabilities of PCDCT scanners. The full spectral capability of PCDCT scanners allows the generation of the entire array of mono-energetic reconstructions, virtual non-contrast (VNC) images, and iodine maps, which have potential advantages in evaluating complex congenital heart disease. For example, following complex congenital cardiac repairs, when distinguishing intraluminal or soft tissue calcifications from contrast, or when evaluating intrastent thrombus. To compare image quality and radiation dose between high-pitch cardiac CT using full spectral PCDCT and dual-source energy-integrating detector CT (EIDCT). This retrospective, IRB-approved study analyzed high-pitch cardiac CTs from January 2021 to October 2023 in pediatric patients (< 18years). Patients were scanned using either PCDCT with full spectral technique ("QuantumPlus") or EIDCT. Radiation doses were measured by CT dose index (CTDI) and dose-length product (DLP). Signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were also calculated, and image quality was assessed using a 5-point Likert scale. Statistical analysis included unpaired T-test, Shapiro-Wilk test, Mann-Whitney test, and kappa coefficients for interrater agreement. Two hundred patients were evaluated, with 100 scanned on PCDCT and 100 on EIDCT. In the PCDCT scanner, 47/100 (47%) were male and 53/100 were female (53%) (P = 0.01). In the EIDCT scanner, 65/100 were male (65%) and 35/100 (35%) were female (P = 0.01). In the PCDCT scanner, 68/100 (68%) (P = 0.05) were ≤ 12months, and 32/100 (32%) (P = 0.05) were > 12months. In the EIDCT, 80/100 (80%) (P = 0.05) were ≤ 12months, and 20/100 (20%) (P = 0.05) were > 12months.In patients ≤ 12months, the CNR were 31.61 in the PCDCT group and 32.14 in the EIDCT group (P = 0.39). For those > 12months, CNR were 30.07 for PCDCT and 25.27 for EIDCT (P = 0.17).In patients ≤ 12months, SNR was significantly lower (P < 0.0001) in PCDCT, compared to EIDCT for the teres minor muscles, while in patients > 12months, SNR was not significantly lower (P = 0.89); SNR was similar between scanners. Radiation doses were significantly higher for PCDCT across both age groups (P < 0.0001). High-pitch cardiac CT with PCDCT using spectral processing resulted in higher radiation doses and lower SNR in infants compared to EIDCT.

  • Research Article
  • 10.1007/s00330-025-12118-7
Comparative analysis of photon-counting and energy-integrating detector CT to identify obstructive coronary artery disease.
  • Nov 14, 2025
  • European radiology
  • Melinda Boussoussou + 17 more

To evaluate the patient-, vessel- and segment-based diagnostic performance of photon-counting detector CT (PCD-CT) compared to energy-integrating detector CT (EID-CT) for detecting ≥ 50% or ≥ 70% stenosis using invasive coronary angiography (ICA) as a reference standard. Patients with stable chest pain and ≥ 50% stenosis detected on dual source PCD-CT who subsequently underwent ICA were prospectively enroled. Diagnostic accuracy was calculated for PCD-CT vs ICA and additionally for a patient cohort scanned with EID-CT with similar risk profiles and disease prevalence. A Monte Carlo simulation based on diagnostic accuracy parameters was performed to estimate the potential reduction in ICA referrals. A total of 143 patients (66 ± 9 years, 27.3% female) with 572 vessels and 2431 segments were evaluated with PCD-CT and ICA. Regarding EID-CT, 109 patients (65 ± 9 years, 31.0% female), 436 vessels and 1853 segments were assessed, with every patient undergoing ICA. PCD-CT demonstrated significantly higher accuracy than EID-CT in detecting ≥ 50% stenosis: 88.1% vs 77.9% (patient level), 91.6% vs 77.8% (vessel level), and 97.7% vs 92.4% (segment level) (p < 0.01 for all). For detecting ≥ 70% stenosis, PCD-CT also showed higher accuracy than EID-CT: 90.9% vs 70.6% (patient level), 94.6% vs 80.9% (vessel level), and 98.6% vs 94.1% (segment level) (p < 0.01 for all). We demonstrated a potential mean reduction of 14.8% in ICA referrals when utilising PCD-CT compared to EID-CT. PCD-CT provides improved per-patient, per-vessel and per-segment diagnostic performance in detecting obstructive CAD in symptomatic patients when compared to patients scanned on EID-CT. PCD-CT may lead to a significant decrease in ICA utilisation. Question Accurate coronary CT angiography guides treatment, but its diagnostic accuracy is limited by various factors. Findings Photon counting detector (PCD)-CT improved diagnostic performance in detecting ≥ 50% or ≥ 70% stenosis, potentially reducing unnecessary ICA referrals by 14.8%. Clinical relevance PCD-CT improves diagnostic accuracy over EID-CT and may reduce unnecessary ICA.

  • Research Article
  • Cite Count Icon 8
  • 10.1007/s00247-024-06078-1
Pediatric contrast-enhanced chest CT on a photon-counting detector CT: radiation dose and image quality compared to energy-integrated detector CT.
  • Oct 28, 2024
  • Pediatric radiology
  • Alexander M El-Ali + 5 more

Photon counting detector (PCD) CT benefits from reduced noise compared with conventional energy-integrating detector (EID) CT, which should translate to improved image quality and reduced radiation exposure for pediatric patients undergoing chest CT with IV contrast. To determine the differences in radiation exposure and image quality of PCD CT and EID CT in pediatric chest CT with intravenous (IV) contrast. In this institutional review board-approved retrospective observational study, 20 scan pairs (20 PCD CT; 20 EID CT) for children who underwent chest CT with IV contrast on both a PCD CT (Siemens NAEOTOM Alpha) and an EID CT (Siemens SOMATOM Definition Edge or Force) within 12months were reviewed independently by three pediatric radiologists for three subjective quality features on 5-point Likert scales: overall quality, small structure delineation, and motion artifact. Objective measures of image quality (image noise, signal-to-noise ratio, and contrast-to-noise ratio) were assessed by a single radiologist in several locations in the chest through region of interest measurement of Hounsfield units (HU) and standard deviation. Patient-related and radiation exposure parameters were collected for each scan and summarized with median and interquartile range (IQR). The Wilcoxon rank-sum test was utilized to compare groups. A P < 0.05 indicated statistical significance. Inter-observer agreement of subjective image quality metrics was analyzed using weighted kappa. Age (14.2years vs 13.8years, P= 0.15), height (P= 0.13), weight (P= 0.21), and BMI (P = 0.24) did not significantly differ between groups. There were 10 male and 3 female patients. Compared to EID CT, PCD CT showed lower radiation exposure parameters including volumetric CT dose index, 1.7mGy (IQR 1.1-2.4mGy) vs 3.8mGy (IQR 2.0-4.7mGy) (P< 0.01), and size-specific dose estimate, 2.6mGy (IQR 1.8-3.1mGy) vs 5.0mGy (IQR 3.3-6.2mGy) (P< 0.01). Objective image quality of lung parenchyma was improved on the PCD CT scanner, including image noise 119.5 HU (IQR 95.4-135.7 HU) vs 143.1 HU (IQR 125.4-169.8 HU) (P < 0.01), signal-to-noise ratio (SNR) -6.1 (IQR -8.4 to -4.8) vs -4.9 (IQR -5.6 to -3.8) (P= 0.01), and contrast-to-noise ratio -63.9 (-84.1 to -57.5) vs -60.5 (-76.3 to -52.5) (P = 0.01). Motion artifact was improved on the PCD CT scanner (P< 0.01). No significant differences in overall image quality or small structure delineation were identified (P= 0.06 and P= 0.31). PCD CT pediatric chest CT had significantly reduced radiation exposure, improved image quality, and reduced motion artifact compared with EID CT.

  • Research Article
  • Cite Count Icon 73
  • 10.1148/ryct.220140
CT Angiography of the Aorta Using Photon-counting Detector CT with Reduced Contrast Media Volume.
  • Feb 1, 2023
  • Radiology: Cardiothoracic Imaging
  • Kai Higashigaito + 10 more

To develop and evaluate a low-volume contrast media protocol for thoracoabdominal CT angiography (CTA) with photon-counting detector (PCD) CT. This prospective study included consecutive participants (April-September 2021) who underwent CTA with PCD CT of the thoracoabdominal aorta and previous CTA with energy-integrating detector (EID) CT at equal radiation doses. In PCD CT, virtual monoenergetic images (VMI) were reconstructed in 5-keV intervals from 40 to 60 keV. Attenuation of the aorta, image noise, and contrast-to-noise ratio (CNR) were measured, and subjective image quality was rated by two independent readers. In the first group of participants, the same contrast media protocol was used for both scans. CNR gain in PCD CT compared with EID CT served as the reference for contrast media volume reduction in the second group. Noninferiority analysis was used to test noninferior image quality of the low-volume contrast media protocol with PCD CT. The study included 100 participants (mean age, 75 years ± 8 [SD]; 83 men). In the first group (n = 40), VMI at 50 keV provided the best trade-off between objective and subjective image quality, achieving 25% higher CNR compared with EID CT. Contrast media volume in the second group (n = 60) was reduced by 25% (52.5 mL). Mean differences in CNR and subjective image quality between EID CT and PCD CT at 50 keV were above the predefined boundaries of noninferiority (-0.54 [95% CI: -1.71, 0.62] and -0.36 [95% CI: -0.41, -0.31], respectively). CTA of the aorta with PCD CT was associated with higher CNR, which was translated into a low-volume contrast media protocol demonstrating noninferior image quality compared with EID CT at the same radiation dose.Keywords: CT Angiography, CT-Spectral, Vascular, Aorta, Contrast Agents-Intravenous, Technology Assessment© RSNA, 2023See also the commentary by Dundas and Leipsic in this issue.

  • Research Article
  • Cite Count Icon 24
  • 10.2214/ajr.23.29950
Photon-Counting Detector CT Versus Energy-Integrating Detector CT of the Lumbar Spine: Comparison of Radiation Dose and Image Quality.
  • Nov 29, 2023
  • AJR. American journal of roentgenology
  • Adrian A Marth + 4 more

BACKGROUND. Photon-counting detector (PCD) CT could be useful to help address the typically high radiation doses of conventional energy-integrating detector (EID) CT of the lumbar spine. OBJECTIVE. The purpose of our study was to compare PCD CT and EID CT of the lumbar spine, both performed using tin filtration, in terms of radiation dose and image quality. METHODS. This study included a prospective sample of 39 patients (22 men, 17 women; mean age, 27.2 years) who underwent investigational PCD CT of the lumbar spine as part of a separate study and a retrospective sample of 39 patients (22 men, 17 women; mean age, 34.9 years) who underwent clinically indicated EID CT of the lumbar spine. In both groups, all examinations were performed using unenhanced technique with tin prefiltration between June 2022 and January 2023. Patients were matched between groups using age, sex, and BMI. A custom gaussian curve-fitting algorithm was used to automatically calculate image noise, SNR, and CNR for each examination, on the basis of all voxels within the image set. Three radiologists independently reviewed examinations to perform a subjective visual assessment of visualization of trabecular architecture, cortical bone, neuroforaminal content, paraspinal muscles, and intervertebral disk, as well as overall image quality, using a 4-point Likert scale (1 = poor, 4 = excellent). PCD CT and EID CT examinations were compared. RESULTS. Mean CTDIvol was 4.4 ± 1.0 (SD) mGy for PCD CT versus 11.1 ± 1.9 mGy for EID CT (p < .001). Mean size-specific dose estimate (SSDE) was 6.2 ± 1.0 (SD) mGy for PCD CT versus 14.2 ± 1.8 mGy for EID CT (p < .001). PCD CT and EID CT examinations were not significantly different in terms of image noise or SNR (both p > .05). PCD CT, in comparison with EID CT, showed significantly higher CNR (mean ± SD, 33.6 ± 3.3 vs 29.3 ± 4.1; p < .001). For all three readers, the median score for overall image quality was 4 (range, 3-4) for both PCD CT and EID CT. PCD CT and EID CT examinations showed no significant difference in terms of any qualitative measure for any reader (all p > .05). CONCLUSION. PCD CT, in comparison with EID CT, yielded significantly lower radiation dose with preserved image quality. CLINICAL IMPACT. The findings support expanded use of PCD CT for lumbar spine evaluation.

  • Research Article
  • Cite Count Icon 5
  • 10.2214/ajr.23.29765
Photon-Counting Detector CT With Denoising for Imaging of the Osseous Pelvis at Low Radiation Doses: A Phantom Study.
  • Nov 22, 2023
  • AJR. American journal of roentgenology
  • Roy P Marcus + 5 more

BACKGROUND. Photon-counting detector (PCD) CT may allow lower radiation doses than used for conventional energy-integrating detector (EID) CT, with preserved image quality. OBJECTIVE. The purpose of this study was to compare PCD CT and EID CT, reconstructed with and without a denoising tool, in terms of image quality of the osseous pelvis in a phantom, with attention to low radiation doses. METHODS. A pelvic phantom comprising human bones in acrylic material mimicking soft tissue underwent PCD CT and EID CT at various tube potentials and radiation doses ranging from 0.05 to 5.00 mGy. Additional denoised reconstructions were generated using a commercial tool. Noise was measured in the acrylic material. Two readers performed independent qualitative assessments that entailed determining the denoised EID CT reconstruction with the lowest acceptable dose and then comparing this reference reconstruction with PCD CT reconstructions without and with denoising, using subjective Likert scales. RESULTS. Noise was lower for PCD CT than for EID CT. For instance, at 0.05 mGy and 100 kV with tin filter, noise was 38.4 HU for PCD CT versus 48.8 HU for EID CT. Denoising further reduced noise; for example, for PCD CT at 100 kV with tin filter at 0.25 mGy, noise was 19.9 HU without denoising versus 9.7 HU with denoising. For both readers, lowest acceptable dose for EID CT was 0.10 mGy (total score, 11 of 15 for both readers). Both readers somewhat agreed that PCD CT without denoising at 0.10 mGy (reflecting reference reconstruction dose) was relatively better than the reference reconstruction in terms of osseous structures, artifacts, and image quality. Both readers also somewhat agreed that denoised PCD CT reconstructions at 0.10 mGy and 0.05 mGy (reflecting matched and lower doses, respectively, with respect to reference reconstruction dose) were relatively better than the reference reconstruction for the image quality measures. CONCLUSION. PCD CT showed better-quality images than EID CT when performed at the lowest acceptable radiation dose for EID CT. PCD CT with denoising yielded better-quality images at a dose lower than lowest acceptable dose for EID CT. CLINICAL IMPACT. PCD CT with denoising could facilitate lower radiation doses for pelvic imaging.

  • Research Article
  • Cite Count Icon 12
  • 10.3174/ajnr.a8350
Neurovascular Imaging with Ultra-High-Resolution Photon-Counting CT: Preliminary Findings on Image-Quality Evaluation.
  • May 17, 2024
  • AJNR. American journal of neuroradiology
  • Adrienn Tóth + 6 more

The first-generation photon-counting detector CT was recently introduced into clinical practice and represents a promising innovation in high-resolution CT imaging. The purpose of this study was to assess the image quality of ultra-high-resolution photon-counting detector CT compared with energy-integrating detector CT and to explore different reconstruction kernel sharpness levels for the evaluation of intracranial aneurysms. Ten patients with intracranial saccular aneurysms who had previously undergone conventional energy-integrating detector CT were prospectively enrolled. CT angiograms were acquired on a clinical dual-source photon-counting detector CT in ultra-high-resolution mode and reconstructed with 4 vascular kernels (Bv36, Bv40, Bv44, Bv48). Quantitative and qualitative image-quality parameters of the intracranial arteries were evaluated. For the quantitative analysis (image noise, SNR, contrast-to-noise ratio), ROIs were manually placed at standard anatomic intracranial and extracranial locations by 1 author. In addition, vessel border sharpness was evaluated quantitatively. For the qualitative analysis, 3 blinded neuroradiologists rated photon-counting detector CT and energy-integrating detector CT image quality for the evaluation of the intracranial vessels (ie, the aneurysms and 9 standard vascular branching locations) on a 5-point Likert-type scale. Additionally, readers independently selected their preferred kernel among the 4 kernels evaluated on photon-counting detector CT. In terms of quantitative image quality, Bv48, the sharpest kernel, yielded increased image noise and decreased SNR and contrast-to-noise ratio parameters compared with Bv36, the smoothest kernel. Compared with energy-integrating detector CT, the Bv48 kernel offered better quantitative image quality for the evaluation of small intracranial vessels (P < .001). Image-quality ratings of the Bv48 were superior to those of the energy-integrating detector CT and not significantly different from ratings of the B44 reconstruction kernel. When comparing side by side all 4 photon-counting detector reconstruction kernels, readers selected the B48 kernel as the best to visualize the aneurysms in 80% of cases. Ultra-high-resolution photon-counting detector CT provides improved image quality for neurovascular imaging. Although the less sharp kernels provided superior SNR and contrast-to-noise ratio, the sharpest kernels delivered the best subjective image quality on photon-counting detector CT for the evaluation of intracranial aneurysms.

  • Research Article
  • Cite Count Icon 132
  • 10.1097/rli.0000000000000816
High-Pitch Photon-Counting Detector Computed Tomography Angiography of the Aorta: Intraindividual Comparison to Energy-Integrating Detector Computed Tomography at Equal Radiation Dose.
  • Aug 6, 2021
  • Investigative Radiology
  • André Euler + 9 more

The aims of this study were to determine the objective and subjective image quality of high-pitch computed tomography (CT) angiography of the aorta in clinical dual-source photon-counting detector CT (PCD-CT) and to compare the image quality to conventional dual-source energy-integrating detector CT (EID-CT) in the same patients at equal radiation dose. Patients with prior CT angiography of the thoracoabdominal aorta acquired on third-generation dual-source EID-CT in the high-pitch mode and with automatic tube voltage selection (ATVS, reference tube voltage 100 kV) were included. Follow-up imaging was performed on a first-generation, clinical dual-source PCD-CT scanner in the high-pitch and multienergy (QuantumPlus) mode at 120 kV using the same contrast media protocol as with EID-CT. Radiation doses between scans were matched by adapting the tube current of PCD-CT. Polychromatic images for both EID-CT and PCD-CT (called T3D) and virtual monoenergetic images at 40, 45, 50, and 55 keV for PCD-CT were reconstructed. Computed tomography attenuation was measured in the aorta; noise was defined as the standard deviation of attenuation; contrast-to-noise ratio (CNR) was calculated. Subjective image quality (noise, vessel attenuation, vessel sharpness, and overall quality) was rated by 2 blinded, independent radiologists. Forty patients were included (mean age, 63 years; 8 women; mean body mass index [BMI], 26 kg/m2). There was no significant difference in BMI, effective diameter, or radiation dose between scans (all P's > 0.05). The ATVS in EID-CT selected 70, 80, 90, 100, 110, and 120 kV in 2, 14, 14, 7, 2, and 1 patients, respectively. Mean CNR was 17 ± 8 for EID-CT and 22 ± 7, 20 ± 6, 18 ± 5, 16 ± 5, and 12 ± 4 for PCD-CT at 40, 45, 50, 55 keV, and T3D, respectively. Contrast-to-noise ratio was significantly higher for 40 and 45 keV of PCD-CT as compared with EID-CT (both P's < 0.05). The linear regression model (adjusted R2, 0.38; P < 0.001) revealed that PCD-CT reconstruction (P < 0.001), BMI group (P = 0.007), and kV of the EID-CT scan (P = 0.01) were significantly associated with CNR difference, with an increase by 34% with PCD-CT for overweight as compared with normal weight patients. Subjective image quality reading revealed slight differences between readers for subjective vessel attenuation and sharpness, whereas subjective noise was rated significantly higher for 40 and 45 keV (P < 0.001) and overall quality similar (P > 0.05) between scans. High-pitch PCD-CT angiography of the aorta with VMI at 40 and 45 keV resulted in significantly increased CNR compared with EID-CT with ATVS at matched radiation dose. The CNR gain of PCD-CT increased in overweight patients. Taking into account the subjective analysis, VMI at 45 to 50 keV is proposed as the best trade-off between objective and subjective image quality.

  • Research Article
  • Cite Count Icon 18
  • 10.1016/j.ejrad.2023.111008
Intra-individual comparison of image quality of the coronary arteries between photon-counting detector and energy-integrating detector CT systems
  • Jul 24, 2023
  • European Journal of Radiology
  • Daniel Pinos + 12 more

Intra-individual comparison of image quality of the coronary arteries between photon-counting detector and energy-integrating detector CT systems

  • Research Article
  • Cite Count Icon 56
  • 10.1148/radiol.222030
Myocardial Characterization with Extracellular Volume Mapping with a First-Generation Photon-counting Detector CT with MRI Reference.
  • Jan 31, 2023
  • Radiology
  • Gilberto J Aquino + 14 more

Background Photon-counting detector (PCD) CT provides comprehensive spectral data with every acquisition, but studies evaluating myocardial extracellular volume (ECV) quantification with use of PCD CT compared with an MRI reference remain lacking. Purpose To compare ECV quantification for myocardial tissue characterization between a first-generation PCD CT system and cardiac MRI. Materials and Methods In this single-center prospective study, adults without contraindication to iodine-based contrast media underwent same-day cardiac PCD CT and MRI with native and postcontrast T1 mapping and late gadolinium enhancement for various clinical indications for cardiac MRI (the reference standard) between July 2021 and January 2022. Global and midventricular ECV were assessed with use of three methods: single-energy PCD CT, dual-energy PCD CT, and MRI T1 mapping. Quantitative comparisons among all techniques were performed. Correlation and reliability between different methods of ECV quantification were assessed with use of the Pearson correlation coefficient (r) and the intraclass correlation coefficient. Results The final sample included 29 study participants (mean age ± SD, 54 years ± 17; 15 men). There was a strong correlation of ECV between dual- and single-energy PCD CT (r = 0.91, P < .001). Radiation dose was 40% lower with dual-energy versus single-energy PCD CT (volume CT dose index, 10.1 mGy vs 16.8 mGy, respectively; P < .001). In comparison with MRI, dual-energy PCD CT showed strong correlation (r = 0.82 and 0.91, both P < .001) and good to excellent reliability (intraclass correlation coefficients, 0.81 and 0.90) for midventricular and global ECV quantification, but it overestimated ECV by approximately 2%. Single-energy PCD CT showed similar relationship with MRI but underestimated ECV by 3%. Conclusion Myocardial tissue characterization with photon-counting detector CT-based quantitative extracellular volume analysis showed a strong correlation to MRI. © RSNA, 2023 Supplemental material is available for this article.

  • Research Article
  • Cite Count Icon 24
  • 10.2214/ajr.23.29077
Comparison of Radiation Dose and Image Quality of Pediatric High-Resolution Chest CT Between Photon-Counting Detector CT and Energy-Integrated Detector CT: A Matched Study.
  • Jul 19, 2023
  • AJR. American journal of roentgenology
  • Marilyn J Siegel + 5 more

BACKGROUND. Photon-counting detector (PCD) CT has been shown to reduce radiation dose and improve image quality in adult chest CT examinations; its potential impact in pediatric CT is not well documented. OBJECTIVE. The purpose of our study was to compare radiation dose, objective image quality, and subjective image quality of PCD CT and energy-integrating detector (EID) CT in children undergoing high-resolution CT (HRCT) of the chest. METHODS. This retrospective study included 27 children (median age, 3.9 years; 10 girls, 17 boys) who underwent PCD CT between March 1, 2022, and August 31, 2022, and 27 children (median age, 4.0 years; 13 girls, 14 boys) who underwent EID CT between August 1, 2021, and January 31, 2022; all examinations comprised clinically indicated chest HRCT. The patients in the two groups were matched by age and water-equivalent diameter. Radiation dose parameters were recorded. One observer placed ROIs to measure objective parameters (lung attenuation, image noise, and SNR). Two radiologists independently assessed subjective measures (overall image quality and motion artifacts) using 5-point Likert scales (1 = highest quality). Groups were compared. RESULTS. PCD CT, in comparison with EID CT, showed lower median CTDIvol (0.41 vs 0.71 mGy, p < .001), DLP (10.2 vs 13.7 mGy × cm, p = .008), size-specific dose estimate (0.82 vs 1.34 mGy, p < .001), and tube current-exposure time product (48.0 vs 202.0 mAs, p < .001). PCD CT and EID CT showed no significant difference in right upper lobe (RUL) lung attenuation (mean, -793 vs -750 HU; p = .09), right lower lobe (RLL) lung attenuation (mean, -745 vs -716 HU; p = .23), RUL image noise (mean, 55 vs 51 HU; p = .27), RLL image noise (mean, 59 vs 57 HU; p = .48), RUL SNR (mean, -14.9 vs -15.8; p = .89), or RLL SNR (mean, -13.1 vs -13.6; p = .79). PCD CT and EID CT showed no significant difference in median overall image quality for reader 1 (1.0 vs 1.0, p = .28) or reader 2 (1.0 vs 1.0, p = .17) or median motion artifacts for reader 1 (1.0 vs 1.0, p = .07) or reader 2 (1.0 vs 1.0, p = .22). CONCLUSION. PCD CT showed significantly reduced dose levels without a significant difference in objective or subjective image quality compared with EID CT. CLINICAL IMPACT. These data expand understanding of the capabilities of PCD CT and support its routine use in children.

  • Research Article
  • 10.1007/s00330-025-12054-6
Reducing contrast media dosage for pulmonary embolism CTPA in PCD-CT: a comparative study of EID-CT and PCD-CT in the era of individualized protocolling.
  • Oct 17, 2025
  • European radiology
  • Lion Stammen + 7 more

To evaluate diagnostic image quality (IQ) for pulmonary embolism detection in photon-counting detector CT (PCD-CT) with significantly reduced contrast media (CM) dose vs conventional energy-integrating detector CT (EID-CT), using optimized and individualized CM protocols. Consecutive CT pulmonary angiography (CTPA) scans performed on the EID-CT (Jan 2024-Mar 2024) with anindividualized kilovoltage (kV) and total body weight (TBW) adapted CM protocol, and on the PCD-CT (Aug 2023-Feb 2024) with a TBW adapted CM protocol matching the 70 kV EID-CT protocol, were retrospectively collected. EID-CT scans were performed at 70-120 kV, based on patients' size, while PCD-CT scans were performed at 120 kV with 55 keV virtual monoenergetic images. Objective IQ assessment included mean attenuation (in Hounsfield Units), signal-to-noise ratio, and contrast-to-noise ratio (CNR). Two board-certified radiologists assessed diagnostic IQ subjectively using a five-point Likert scale. In 140 EID-CT and 118 PCD-CT scans, PCD-CT reduced total iodine load by 26.7% and CT dose index volume by 24.4%. Objective IQ parameters showed no significant differences, except for a decrease in CNR in the proximal pulmonary arteries in PCD-CT scans (p = 0.02). Subjective IQ was rated as moderate/good by observers 1 and 2 in 94.9% and 97.9% of the EID-CT scans, respectively, and 96.5% and 100% of the PCD-CT scans, respectively. PCD-CT with a TBW-adapted CM protocol can achieve substantial reductions in both CM and radiation dose compared to EID-CT with individualized kV and TBW adapted CM protocols, while maintaining diagnostic IQ in CTPA scans. Question The feasibility of CM reduction in PCD-CT vs EID-CT with highly individualized CM protocols, while maintaining diagnostic IQ. Findings PCD-CT reduces total iodine load by 26.7% and CT dose index volume by 24.4% compared to EID-CT, with comparable objective and subjective diagnostic IQ. Clinical relevance This approach enables CM reduction, potentially lowering patient risk and providing environmental and financial benefits. Additionally, PCD-CT allows for CM and radiation dose reduction while maintaining comparable IQ.

  • PDF Download Icon
  • Research Article
  • Cite Count Icon 106
  • 10.3174/ajnr.a5402
Photon-Counting CT of the Brain: In Vivo Human Results and Image-Quality Assessment.
  • Oct 5, 2017
  • American Journal of Neuroradiology
  • A Pourmorteza + 7 more

Photon-counting detectors offer the potential for improved image quality for brain CT but have not yet been evaluated in vivo. The purpose of this study was to compare photon-counting detector CT with conventional energy-integrating detector CT for human brains. Radiation dose-matched energy-integrating detector and photon-counting detector head CT scans were acquired with standardized protocols (tube voltage/current, 120 kV(peak)/370 mAs) in both an anthropomorphic head phantom and 21 human asymptomatic volunteers (mean age, 58.9 ± 8.5 years). Photon-counting detector thresholds were 22 and 52 keV (low-energy bin, 22-52 keV; high-energy bin, 52-120 keV). Image noise, gray matter, and white matter signal-to-noise ratios and GM-WM contrast and contrast-to-noise ratios were measured. Image quality was scored by 2 neuroradiologists blinded to the CT detector type. Reproducibility was assessed with the intraclass correlation coefficient. Energy-integrating detector and photon-counting detector CT images were compared using a paired t test and the Wilcoxon signed rank test. Photon-counting detector CT images received higher reader scores for GM-WM differentiation with lower image noise (all P < .001). Intrareader and interreader reproducibility was excellent (intraclass correlation coefficient, ≥0.86 and 0.79, respectively). Quantitative analysis showed 12.8%-20.6% less image noise for photon-counting detector CT. The SNR of photon-counting detector CT was 19.0%-20.0% higher than of energy-integrating detector CT for GM and WM. The contrast-to-noise ratio of photon-counting detector CT was 15.7% higher for GM-WM contrast and 33.3% higher for GM-WM contrast-to-noise ratio. Photon-counting detector brain CT scans demonstrated greater gray-white matter contrast compared with conventional CT. This was due to both higher soft-tissue contrast and lower image noise for photon-counting CT.

Save Icon
Up Arrow
Open/Close
  • Ask R Discovery Star icon
  • Chat PDF Star icon

AI summaries and top papers from 250M+ research sources.