Abstract

The aim of our work was to evaluate the analytical performance of the anti-cyclic citrullinated peptide determination by a two-step immunoassay using microparticle chemiluminescence immunoassay (CMIA) technology, in accordance with the Scope A criteria of the guide of the verification/validation of medical biology methods. We evaluated the repeatability and the intermediate precision of the assay. The results obtained are very satisfactory for the two levels (low and high) both for intermediate fidelity, with coefficients of variation (CV) of CV1= 1.9% and CV2=1.8% respectively, and for repeatability, with coefficients of variation of CV1 = 1.7%, CV2= 1.61%. respectively. The results obtained made it possible to verify the method's performance and compare it with the analytical objectives set in order to meet the regulatory and normative requirements set by the supplier and learned societies.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.