Abstract

A 79-year-old woman with severe mitral annular calcification was scheduled for mitral valve replacement. A SAPIEN 3 valve was implanted in mitral position using an open surgical approach. Immediately after cardiopulmonary bypass, bleeding from an unidentified source was encountered. Cardiopulmonary bypass was emergently resumed and a laceration of the left ventricular apex due to the valve delivery system was detected. Risk factors specific to the open surgical approach include a decompressed ventricle, decreased annulus to apical distance, and the absence of continuous fluoroscopic and echocardiographic imaging. These create a clinical scenario where risk of ventricular perforation is increased compared with traditional intravascular transcatheter valve delivery.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.