Abstract

Profiles of thermal flux, epithermal flux, f and α have been determined for a multisample irradiation container in use with k0-based NAA. It could be concluded that for accurate NAA with the IRI-reactor, samples should be surrounded by at least two thermal flux monitors. Since f and α values were found to vary considerably between irradiations as compared to the variations within one container, the possibility of determining f and α just once for the irradiation facility used, and monitoring only the thermal flux in future irradiations, was ruled out. The use of average values for f and α within a container was shown to yield unacceptable inaccuracies mainly for Zr and U determinations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.