Abstract

OBJECTIVETo characterize glucose levels during daily living using continuous glucose monitors (CGMs) in nondiabetic individuals.RESEARCH DESIGN AND METHODSSeventy-four healthy children, adolescents, and adults aged 9–65 years with normal glucose tolerance used a blinded CGM device for 3 to 7 days.RESULTSSensor glucose concentrations were 71–120 mg/dl for 91% of the day. Sensor values were ≤60 or >140 mg/dl for only 0.2% and 0.4% of the day, respectively. Sensor glucose concentrations were slightly higher in children than adults (P = 0.009) and were slightly lower during the night than day (95 vs. 99 mg/dl, P < 0.001).CONCLUSIONSGlucose values ≤60 and >140 mg/dl, measured with CGM, are uncommon in healthy, nondiabetic individuals. CGM may be useful to evaluate glucose tolerance in nondiabetic individuals over time. Furthermore, these data provide a basis for comparison for studies that use CGM to assess glucose control in subjects with diabetes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.