Abstract

The purpose of the study was to determine the variability of measurement that might occur with rotation of the scapula in the coronal plane when computed tomography scanning is used. Ten dry scapula specimens had 4 sets of computed tomography scans performed at +20°, +10°, neutral, and −10° of scapula rotation. The angle of glenoid version was then measured according to the technique of Friedman. The measured value for glenoid version was noted to vary by as much as 10° on the same specimen with minor rotation of the scapula. This result occurred because of differences in the curvature of the vertebral border of the scapula. The measured angle of glenoid version was most likely to vary with rotation of the scapula 10° downward (ie, glenoid articular surface facing 10° upward). For accurate and reproducible measurement of the glenoid version, it is essential that in the scout view the glenoid orientation is neutral (ie, glenoid surface is perpendicular to the plane of the computed tomography cut).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.