Abstract

The 2-min protocol (1 + 1) for the βeta-s.t.a.r. (manufactured by Neogen Corporation, Lansing, MI, USA) was validated at the Technology and Food Science Unit of the Institute for Agricultural and Fisheries Research according to Commission Decision 2002/657/EC. The test was very selective for the group of β-lactam compounds: the only interference found was by clavulanic acid at 2500 µg kg−1 and above. The modified protocol (βeta-s.t.a.r. 1 + 1) detected all β-lactams with a maximum residue limit (MRL) in milk, but not all these compounds were detected at their respective MRL. The detection of cefalexin (detection capability = 6000 µg kg−1; MRL = 100 µg kg−1) and penethamate (detection capability = 80 µg kg−1; MRL = 4 µg kg−1) was especially poor, and also ceftiofur was only detected from 500 µg kg−1 (MRL = 100 µg kg−1). The repeatability of the reader and of the test was very good. The test was very robust: test results were not significantly influenced by small changes in the test protocol, by the milk composition or by the type of milk. The test was also suitable to test the milk of animal species other than cow. Favourable results were obtained in testing monitoring samples, in two national ring trials, and in an international proficiency test. The βeta-s.t.a.r. 1 + 1 is a very fast, simple, and reliable test that could be used at the farm level to prevent tanker milk contamination by β-lactams.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.