Abstract

The objective of this study was to build and validate the FE models of thoracolumbar junctional T11-T12 and T12-L1 functional spinal units (FSUs) and compare the biomechanical responses of the two FSUs under physiological loading modes: flexion, extension, lateral bending and axial rotation. Anatomically accurate FE models of thoracolumbar T11-T12 and T12-L1 FSUs were developed and validated against published experimental results in terms of load displacement responses and range of motion (ROM) under flexion and extension pure moments of 7.5 Nm, left and right lateral bending pure moments of 7.5 Nm and left and right axial torque of 7.5 Nm. The overall responses predicted by the T11-T12 and T12-L1 FE models showed differences in stiffness under different load configurations. Amongst all loading configurations, the motions at T11-T12 and T12-L1 were the stiffest under axial torque. The lateral bending motions of T11-T12 and T12-L1 were relatively flexible. Under sagittal moments, the motion in extension was greater than in flexion at level T11-T12, while the rotation in flexion was greater than in extension at level T12-L1.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.