Abstract

A novel spine test machine was developed for physiological loading of spinal segments. It can be used in conjunction with external motion-capture systems (EMCS) to measure angular displacement, but can also measure in-plane rotations directly, though the inherent error is unknown. This study quantified error inherent in the displacement measurement of the machine. Synthetic specimens representative of cadaveric spinal specimens were tested. Machine displacement was compared to EMCS displacement. The maximum machine displacement error was <2 deg for lumbar and thoracic specimens. The authors suggest that researchers use EMCS in conjunction with the test machine when high accuracy measurements are required.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.