Abstract
The objective of the study was to independently validate a calibrated commercial handheld near infrared (NIR) spectroscopic device and test its repeatability over time using phenotypically diverse populations of Australian lamb. Validation testing in eight separate data sub-groups (n = 1591 carcasses overall) demonstrated that the NIR device had moderate precision (R2 = 0.4–0.64, RMSEP = 0.70–1.22%) but fluctuated in accuracy between experimental site demonstrated by variable slopes (0.50–0.94) and biases (−0.86–0.02). The repeatability experiment (n = 10 carcasses) showed that time to scan post quartering affected NIR measurement from 0 to 24 h (P < 0.001). On average, NIR IMF% was 0.97% lower (P < 0.001) at 24 h (4.01% ± 0.166), compared to 0 h. There was no difference (P > 0.05) between Time 0 and 1 h or Time 0 and 4 h or between replicate scans within each time point. This study demonstrated the SOMA NIR device could predict lamb chemical IMF% with moderate precision and accuracy, however additional work is required to understand how loin preparation, blooming and surface hydration affect NIR measurement.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.