Abstract

The objective of the study was to independently validate a calibrated commercial handheld near infrared (NIR) spectroscopic device and test its repeatability over time using phenotypically diverse populations of Australian lamb. Validation testing in eight separate data sub-groups (n = 1591 carcasses overall) demonstrated that the NIR device had moderate precision (R2 = 0.4–0.64, RMSEP = 0.70–1.22%) but fluctuated in accuracy between experimental site demonstrated by variable slopes (0.50–0.94) and biases (−0.86–0.02). The repeatability experiment (n = 10 carcasses) showed that time to scan post quartering affected NIR measurement from 0 to 24 h (P < 0.001). On average, NIR IMF% was 0.97% lower (P < 0.001) at 24 h (4.01% ± 0.166), compared to 0 h. There was no difference (P > 0.05) between Time 0 and 1 h or Time 0 and 4 h or between replicate scans within each time point. This study demonstrated the SOMA NIR device could predict lamb chemical IMF% with moderate precision and accuracy, however additional work is required to understand how loin preparation, blooming and surface hydration affect NIR measurement.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.