Abstract

Abstract In this study, we introduce the current situation of the membrane inspection industry. We have developed a new integrated rapid membrane inspection machine (MIM) to test the performances of nanofiltration or ultrafiltration membranes by measuring the silt density index (SDI) value (i.e., the key performance indicator) of the water filtered. This study optimizes many key conditions include temperature, pH value and pressures, then establishes a new SDI detection method of ultrafiltration membranes by using MIM. We set the temperature of the test liquid to 25 °C, pH in the range of 7–8, and pressure to approximately 0.10 MPa. The results reflect that MIM can be used to effectively determine the SDI of ultrafiltration membranes using the new method and the results of the method validation are satisfactory.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.