Abstract

The purpose of this work was to assess the stability of fiducial markers in the prostate bed and compared their use to surgical clips. In this study, 3-4 gold fiducial markers were transrectally implanted in the prostate bed of 14 patients. The stability of the fiducial markers position (fiducial markers fixity) over an EBRT course was assessed. Furthermore, the advantages of the fiducial markers compared to the surgical clips were assessed and the interobserver variation between the two technologies was compared. The mean fiducial marker migration during a course of EBRT was small with 1.2 mm (SD ± 0.8 mm). Compared to fiducial markers, the matches with surgical clips were mismatched ≥ 2 mm in 68% of treatments. This discrepancy of > 2 mm was on average 3.7 ± 1.3 mm. There was less interobserver variability for matching of fiducial markers (0.8 ± 0.7 mm) than for surgical clips (2.0 ± 1.6 mm). Fiducial markers showed less interobserver variability in matching and less variation in position than surgical clips. Fiducial markers could ultimately help in reducing treatment margins.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.