Abstract

ObjectivesWe developed and validated the use of synthetic urine as a matrix substitute for standard and quality control material preparation in the clinical assessment of iodine status in urine. Design and methodsMeasurement of iodine in urine was conducted using inductively coupled-plasma mass spectrometry. Analytical and clinical recoveries were assessed to investigate comparability between synthetic urine and pooled patient urine. Method performance characteristics were determined in accordance with clinical laboratory standards. ResultsEstablished assay performance characteristics included inter- and intra-assay imprecision <10%, carryover of <0.2%, analytical measurement range of 5 to 1000μg/L, limit of quantification of 5μg/L (coefficient of variation <10%), proportional bias of 0.92 and constant bias of 8.8 in comparison to an outside reference laboratory. ConclusionsSynthetic urine is an appropriate alternative matrix for standard and quality control material preparation for the measurement of iodine in urine.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.