Abstract

The leakage neutron spectrum measurements have been done on benchmark spherical assembly-nickel sphere with a diameter of 50 cm and a copper cube (block) with dimensions of 49.5 x 49.5 x 48 cm3 in Research Centre Rez (RC Rez). The 252Cf neutron source was placed into the centre of nickel sphere and copper cube. The proton recoil method was used for the neutron spectrum measurement using spherical hydrogen proportional detectors (HPD) with pressure of 400 and 1000 kPa (diameter of detectors is 4 cm) and scintillation stilbene (ST) detector (diam. of 1 x 1 cm). The neutron energy range of spectrometer is from 0.04 MeV to 1.3 MeV for HPD and from 1 MeV to 12 MeV for ST. The adequate MCNP neutron spectrum calculations based on data libraries ENDF/B-VII.1, ENDF/B-VIII.0, JEFF-3.3, BROND-3.1 were done and compared with the experiment, i.e., calculation to experiment ratio C/E was determined.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.