Abstract

A simple colorimetric flow-injection system for the determination of Cu(II) based on a complexation reaction with nitroso-R salt is described. The chemical and FIA variables were established using the univariate and simplex methods. A small volume of Cu(II) was mixed with merged streams of nitroso-R salt and acetate buffer solutions. The absorbance of the complex was continuously monitored at 492 nm. The calibration curve over the concentration range 1.0-7.0 microg ml(-1) was obtained. The relative standard deviation for determining 4.0 microg ml(-1) Cu(II) was 0.47% (n = 11). The detection limit (3sigma) was 0.68 microg ml(-1) and the sample throughput was 150 h(-1). The validity of the method has been satisfactorily examined for the determination of Cu(II) in wastewater and copper ore samples. The accuracy was found to be high, because the student t-values were calculated to be less than the theoretical values when the results were compared with those obtained by FAAS.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.