Abstract

Low noise underground environments offer conditions allowing assessment of ultimate performance of high sensitivity sensors such as accelerometers, gyrometers, seismometers⋯ Such facilities are for instance ideal for observing the tiny signals of interest for geophysical studies. Laboratoire Souterrain a Bas Bruit (LSBB) in which we have installed our cold atom gravimeter provides such an environment. We report here the best short term sensitivity ever obtained without any ground vibration isolation system with such an instrument: 10−8 m s−2 in 100 s measurement time.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.