Abstract
This paper presents recent progress in reducing the measurement uncertainty for crystalline silicon (c-Si) and thin-film PV modules. It describes the measurement procedure and the uncertainty analysis, as applied at the CalLab PV Modules, Fraunhofer ISE's laboratory for module measurements. The uncertainty analysis covers the complete calibration process in detail, including measurements, correction to standard testing conditions, and determination of electrical module parameters ( <i xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">I</i> <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">SC</sub> , <i xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">P</i> <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">MPP</sub> , <i xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">V</i> <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">OC</sub> , etc.) from the I-V curve. Differences between c-Si and thin-film modules are addressed, most importantly in terms of spectral mismatch factor and short timescale stability problems. The paper outlines the importance of a comprehensive quality assurance system in a calibration laboratory as a prerequisite for accurate measurements on a daily basis. Particular attention is paid to results from a series of measurements taken every three weeks over a three-year period and conducted as part of the quality assurance system. In conclusion, this paper introduces a best-case uncertainty for c-Si module calibration of 1.6% for <i xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">P</i> <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">MPP</sub> and 1.3% for <i xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">I</i> <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">SC</sub> . This represents the lowest reported uncertainty for full size module calibration in a laboratory so far. The presented uncertainty in <i xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">P</i> <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">MPP</sub> of cadmium telluride and single-junction amorphous silicon modules is 2.9% and 1.8%, respectively. All mentioned uncertainties are expanded uncertainties ( <i xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">k</i> = 2).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.