Abstract

Skeletal muscle is considered as a near-constant volume system, and the contractions of the muscle are related to the changes in tissue thickness. Assessment of the skeletal muscle contractile parameters such as maximum contraction thickness (), contraction time (), contraction velocity (), sustain time (), and half-relaxation () provides valuable information for various medical applications. This paper presents a single-element wearable ultrasonic sensor (WUS) and a method to measure the skeletal muscle contractile parameters in A-mode ultrasonic data acquisition. The developed WUS was made of double-layer polyvinylidene fluoride (PVDF) piezoelectric polymer films with a simple and low-cost fabrication process. A flexible, lightweight, thin, and small size WUS would provide a secure attachment to the skin surface without affecting the muscle contraction dynamics of interest. The developed WUS was employed to monitor the contractions of gastrocnemius (GC) muscle of a human subject. The GC muscle contractions were evoked by the electrical muscle stimulation (EMS) at varying EMS frequencies from 2 Hz up to 30 Hz. The tissue thickness changes due to the muscle contractions were measured by utilizing a time-of-flight method in the ultrasonic through-transmission mode. The developed WUS demonstrated the capability to monitor the tissue thickness changes during the unfused and fused tetanic contractions. The tetanic progression level was quantitatively assessed using the parameter of the fusion index (FI) obtained. In addition, the contractile parameters (, , , , and ) were successfully extracted from the measured tissue thickness changes. In addition, the unfused and fused tetanus frequencies were estimated from the obtained FI-EMS frequency curve. The WUS and ultrasonic method proposed in this study could be a valuable tool for inexpensive, non-invasive, and continuous monitoring of the skeletal muscle contractile properties.

Highlights

  • The analysis of the skeletal muscle mechanical characteristics is of great interest for a wide range of medical applications

  • The periodical change of the tissue thickness corresponding to electrical muscle stimulation (EMS) frequency was clearly observed from the GC muscle contraction at the EMS frequency from 2 to 12 Hz

  • The tissue thickness changes due to GC muscle contractions were successfully obtained at each EMS frequency by utilizing the ultrasonic TT mode in A-mode data acquisition, where two wearable ultrasonic sensor (WUS) employed as transmitter and receiver

Read more

Summary

Introduction

The analysis of the skeletal muscle mechanical characteristics is of great interest for a wide range of medical applications. The measurement of changes in the muscle physical parameters during contractions provides valuable information. Monitoring of the change in muscle thickness during contractions would be a useful tool to assess the muscle contractile properties [9,15]. Several studies have provided evidence that the relative changes of the muscle contractile parameters were associated with the skeletal muscle atrophy [16,17], the percentage of muscle fiber types [18,19], general and local muscle fatigue [9,20,21,22,23,24,25], and muscle force production [26,27,28]. The common non-invasive methods include surface electromyography (SEMG), mechanomyography (MMG), and ultrasound imaging

Objectives
Methods
Results
Conclusion

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.